Liquid Chromatograph with Single Quadrupole Mass Spectrometer
ACQUITY Arc System
ACQUITY QDa Mass Detector
The ACQUITY Arc System is a quaternary-based LC system designed to achieve both HPLC and UHPLC separations in one instrument. The system allows to easily switch between HPLC and UHPLC operation with multi-flow technology to select the dwell volume that fits best the column separation.
- Maximum operating pressure: 9500 psi.
- pH range: 1–12.5.
- Operating flow rate range: up to 5 mL/min.
- Sample compartment: 4–40 °C.
- Pump options: quaternary.
- Column capacity: up to 15 columns of 150 mm length and up to 4.6 mm inner diameter.
- Maximum sample capacity: up to 20 plates (with Sample Organizer).
ACQUITY QDa is is a mass detector designed for analytical scientists who need
mass spectral data without the complexity of a mass spectrometer, to confirm compound identity and quantify compounds with no UV response or at levels not accessible by optical detection. With the on/off simplicity, the ACQUITY QDa Detector fully automates sample analysis and eliminates sample-specific or user adjustments typical of traditional mass spectrometers.
- Acquisition modes: Full scan MS; Selected Ion Recording (SIR).
- Mass range: 30 to 1250 m/z.
- Scan speed: automatically-optimized for enhanced data quality for acquisition rates of up to e.g. 10 Hz for m/z 100 to 1000 or 20 Hz for m/z 50 to 500.
- Mass accuracy: better than ±0.2 Da over full mass range.
- Mass stability: Mass drift less than 0.1 Da over a 24 hour period.
- Mass resolution: Automated mass resolution control (0.7 Da) for constant data quality.
📍 The instrument is located at DIFARMA of the
University of Salerno.
Go Back